1. પરિચય
The GIMA OXY 6 Finger Pulse Oximeter is a portable device designed for non-invasive spot-checking of arterial oxygen saturation (SpO2) and pulse rate. It is suitable for use in medical and hospital environments, as well as for personal use under medical guidance. This manual provides essential information for the safe and effective operation of the device.
2. સલામતી માહિતી
Please read all safety warnings and precautions before using the device. Failure to do so may result in inaccurate readings or device malfunction.
- This device is intended for spot-checking and not for continuous monitoring.
- MRI અથવા CT વાતાવરણમાં ઓક્સિમીટરનો ઉપયોગ કરશો નહીં.
- ઉપકરણને પાણી અને ભેજથી દૂર રાખો.
- Consult a medical professional for interpretation of readings. This device is not a substitute for professional medical advice.
- Ensure the finger is clean and free of nail polish or artificial nails, as these can affect accuracy.
- Avoid excessive movement during measurement to ensure stable readings.
3. ઉત્પાદન ઓવરview
The GIMA OXY 6 features a high-resolution OLED color display showing SpO2, pulse rate, pulse bar graph, perfusion index (PI), and battery level. It is designed for ease of use with a compact form factor.

આકૃતિ 1: આગળ view of the GIMA OXY 6 Pulse Oximeter, displaying SpO2 and pulse rate values on its screen.
3.1. પેકેજ સામગ્રી

આકૃતિ 2: The package includes the GIMA OXY 6 Pulse Oximeter, a plastic case, AAA batteries, and a lanyard.
- GIMA OXY 6 Pulse Oximeter
- 2 x AAA બેટરી
- પ્લાસ્ટિક વહન કેસ
- લેનયાર્ડ
- સૂચના માર્ગદર્શિકા

આકૃતિ 3: The GIMA OXY 6 Pulse Oximeter attached to a lanyard, demonstrating its portability and ease of carrying.
4. સેટઅપ
4.1. બેટરી ઇન્સ્ટોલેશન
- ઉપકરણની પાછળ સ્થિત બેટરી કમ્પાર્ટમેન્ટ કવર ખોલો.
- Insert two AAA batteries according to the polarity indicators (+/-) inside the compartment.
- બેટરી કમ્પાર્ટમેન્ટ કવર સુરક્ષિત રીતે બંધ કરો.

આકૃતિ 4: Open battery compartment showing placement for two AAA batteries.
5. ઓપરેટિંગ સૂચનાઓ
5.1. માપ લેવું
- Ensure the device has fresh batteries installed.
- cl ખોલોamp ઓક્સિમીટરનું.
- Insert one finger (preferably the index, middle, or ring finger) into the rubber opening of the oximeter, ensuring the finger is fully inserted and resting on the sensor. The recommended finger thickness for this model is 1-22 mm.
- સીએલ છોડોamp. The device will automatically turn on and begin measurement.
- Keep your hand still during the measurement. Avoid bending your finger or making hand movements.
- Readings for SpO2, Pulse Rate, and Perfusion Index (PI) will appear on the display within a few seconds.

આકૃતિ 5: The GIMA OXY 6 Pulse Oximeter correctly placed on a finger for measurement.
5.2. ડિસ્પ્લેને સમજવું
- SpO2: Peripheral Oxygen Saturation, displayed as a percentage.
- PR: Pulse Rate, displayed in beats per minute (bpm).
- પીઆઈ: Perfusion Index, a numerical value indicating the strength of the pulse signal at the sensor site.
- પલ્સ બાર ગ્રાફ: Visual representation of pulse strength.
- બેટરી સૂચક: વર્તમાન બેટરી સ્તર બતાવે છે.
5.3. Automatic Power Off
The device will automatically power off after approximately 8 seconds of inactivity or when the finger is removed.
5.4. લો વોલ્યુમtage એલાર્મ
જ્યારે બેટરી વોલtage is low, a low voltage alarm will be indicated on the display. Replace the batteries promptly to ensure accurate operation.
5.5. વૉઇસ ગાઇડ
The GIMA OXY 6 features an 8-language voice guide (English, French, Italian, Spanish, Greek, Portuguese, Arabic) for enhanced usability. Refer to the full manual for specific instructions on activating and changing voice guide settings.
6. જાળવણી
6.1. સફાઈ
- Clean the surface of the oximeter with a soft cloth dampened with medical alcohol (70% ethanol or isopropanol).
- ઉપકરણને પાણી કે કોઈપણ સફાઈ દ્રાવણમાં બોળશો નહીં.
- Ensure the device is completely dry before reuse.
6.2. બેટરી રિપ્લેસમેન્ટ
જ્યારે ઓછી વોલ્યુમ હોય ત્યારે બેટરી બદલોtage indicator appears. Always use two new AAA alkaline batteries. Dispose of used batteries according to local regulations.
6.3. સંગ્રહ
Store the oximeter in its plastic carrying case in a cool, dry place, away from direct sunlight and extreme temperatures. Remove batteries if the device will not be used for an extended period.
7. મુશ્કેલીનિવારણ
| સમસ્યા | સંભવિત કારણ | ઉકેલ |
|---|---|---|
| પાવર ચાલુ થયા પછી કોઈ પ્રદર્શન નથી | બેટરીઓ મૃત અથવા ખોટી રીતે ઇન્સ્ટોલ કરેલી છે. | ધ્રુવીયતાનું અવલોકન કરીને, બેટરીઓ બદલો અથવા તેમને યોગ્ય રીતે ફરીથી ઇન્સ્ટોલ કરો. |
| અચોક્કસ વાંચન | Finger not properly inserted; excessive movement; nail polish; cold finger. | Re-insert finger correctly; remain still; remove nail polish; warm the finger. |
| "Finger Out" or similar error message | Finger not detected or improperly placed. | Ensure finger is fully inserted and resting on the sensor. |
| ઉપકરણ અનપેક્ષિત રીતે બંધ થાય છે | Low battery; automatic power-off due to inactivity. | Replace batteries; ensure finger is inserted for measurement. |
8. સ્પષ્ટીકરણો
| મોડલ નંબર | 34285 (OXY 6) |
| ડિસ્પ્લે પ્રકાર | OLED Color Display |
| માપન પરિમાણો | SpO2, Pulse Rate, Perfusion Index (PI) |
| SpO2 Measurement Range | (Typical range, refer to device for exact range) |
| પલ્સ રેટ મેઝરમેન્ટ રેન્જ | (Typical range, refer to device for exact range) |
| ચોકસાઈ | High precision and repeatability with noise suppression |
| પાવર જરૂરીયાતો | 2 x AAA 1.5V Alkaline Batteries |
| બેટરી જીવન | Approximately 3 hours of continuous operation with full batteries |
| પરિમાણો (L x W x H) | 16 x 5.5 x 10 સેમી |
| વજન | 200 ગ્રામ (બેટરી સહિત) |
| Finger Thickness Range | 1-22 મીમી |
| વૉઇસ માર્ગદર્શિકા | 8 languages (GB, FR, IT, ES, GR, PT, Arabic) |
9. વોરંટી અને સપોર્ટ
Information regarding specific warranty terms and customer support for the GIMA OXY 6 Pulse Oximeter is not provided in this document. Please refer to the product packaging or contact GIMA customer service for details.