ReLeaf RESTORATIVE Basic Setup Kit

ReLeaf-RESTORATIVE-Basic-Setup-Kit-product

વિશિષ્ટતાઓ

  • Material: Soft, flexible thermoplastic elastomer
  • 100% BPA and latex-free
  • Intended Use: Restorative dental procedures

ઉત્પાદન વર્ણન/ઉદ્દેશિત ઉપયોગ

ReLeaf® Restorative is a hands-free, high-volume dental suction device that assists practitioners with evacuation, isolation, retraction, and maintaining a dry field during general dentistry, restorative, and hygiene procedures. ReLeaf® is for clinical use by dental professionals.
The Leaf mouthpiece comprises a soft, flexible thermoplastic elastomer that is 100% BPA and latex-free. No health hazards have been identified with the use of this material.
The ReLeaf® hoses comprise of a silicone rubber (silicone and silicon dioxide).

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (1)

સ્થાપન અને દર્દીનો ઉપયોગ

  1. Fold the two Leaf halves together by pushing together the small protruding knob into the opposing hole at the Restorative Leaf base. Insert the larger knob into the opposing hole at the Leaf tip. Squeeze the two halves of the Leaf together to ensure each side securely attaches to the other. Attach the Leaf to the Restorative Adapter on the Short Hose (8” Quick Disconnect Hose) with the smaller leaf face facing away from the hose. Ensure the Leaf is securely attached.
  2. Insert the ReLeaf® Bite Block onto the peg closest to the Leaf on the Restorative Adapter. Making sure the concaved slope is in contact with the Restorative Leaf.ReLeaf-RESTORATIVE-Basic-Setup-Kit- (2)
  3. Attach the free end of the
    Short Hose to the Quick Disconnect Adapter on the Long Hose (48” HVE Attachment Hose). Insert the coupler at the other end of the Long Hose securely into an open HVE port. It may be necessary to wet the
    O- ring inside the unused port so the Adapter can slide in easily.
  4. Place the ReLeaf® Bite Block on the opposing arch you are working on. While having the Leaf sweep in front of the tongue and wrap around the retromolar area to the opposing cheek side.
  5. When transitioning between patients, a cleaned and sterilized Bite Block, Restorative Adapter, and Short Hose must be utilized before attaching a new Leaf.ReLeaf-RESTORATIVE-Basic-Setup-Kit- (3)

વિરોધાભાસ
કોઈ જાણીતા વિરોધાભાસ નથી.

સાવચેતીનાં પગલાં

Risk of allergic reactions cannot be excluded for those predisposed to allergies.

STERILIZATION, DISPOSAL AND STORAGE

All gray and white components of the ReLeaf™ system are compatible with standard dental sterilization protocols. These parts can be: wiped with dental-approved disinfectant wipes or sprays, run through dental atomizer, processed through an ultrasonic bath, run through a thermal washer, and run through autoclave cycle per sterilization cycle table.

  • Leaves: One Time Use, Disposable
    The disposable Leaves are meant for single use only. Due to the nature of the material and design of the Leaves, they should not be sterilized and reused. Leaf mouthpieces must be properly discarded in a clinical waste container to prevent transmission of infectious pathogens.
  • Bite Block, Restorative Adapter and Short Hose
    100% Autoclavable and Reusable.
    Before placement in an ultrasonic cleaner, rinse the Bite Block, Restorative Adapter and Short Hose with water and brush both external and lumen surfaces with a soft bristle brush for the exterior surface and a soft bristle lumen brush or pipe cleaner for the interior surfaces. Place in an ultrasonic unit filled with enzymatic cleaner. (note: follow instructions with enzymatic cleaner for proper mixing, exposure time, etc., per manufacturer instructions). Rinse with water, place in an autoclavable pouch. Run product through a wrapped / pouched cycle in the autoclave.
  • Long Hose, Quick Disconnect Adapter and HVE Adapter
    100% Autoclavable and Reusable.
    The Long Hose should stay connected to the HVE until ready for routine evacuation cleaning with an enzymatic evacuation cleaner. Running evacuation cleaners through the hose to clean out any debris at least once per week is recommended. It is not necessary to run evacuation cleaners through the hose between each patient. If running through the autoclave, follow instructions listed above for Bite Block, Restorative Adapter & Short Hose.

સંગ્રહ
Please ensure Releaf® is stored at room temperature.
The Long Hose and Short Hose are designed to be held in the HVE Adapter when unused. (See image)

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (4)

માન્ય નસબંધી ચક્રો

ગુરુત્વાકર્ષણ વિસ્થાપન ઓટોક્લેવ

  • તાપમાન: ૧૨૧°C/૨૫૦°F
  • MINIMUM EXPOSURE TIME: 30 minutes
  • ન્યૂનતમ સૂકવવાનો સમય: ૩૦ મિનિટ
માન્ય નસબંધી ચક્રો
ગુરુત્વાકર્ષણ વિસ્થાપન ઓટોક્લેવ પ્રી-વેક્યુમ ઓટોક્લેવ
TEMPERATURE: 121°સી/250°F TEMPERATURE: 132°સી/270°F
MINIMUM EXPOSURE TIME: 30 minutes MINIMUM EXPOSURE TIME: 4 minutes
ન્યૂનતમ સૂકવવાનો સમય: ૩૦ મિનિટ ન્યૂનતમ સૂકવવાનો સમય: 20 મિનિટ

મુશ્કેલી નિવારણ

Proper Releaf® Restorative Position
Ensure the long end of the Restorative Leaf wraps around the retromolar area on the quadrant you are working on. To assist in evacuation, when a small amount of water and saliva build up behind the Restorative Leaf, have the patient slowly push their tongue against the Leaf while closing down on the Bite Block.
This will evacuate any remaining water that flows behind the Leaf.

છૂટક જોડાણ
If you find the coupler becoming loose, it may be necessary to replace the O-ring.

Releaf® is a registered trademark of Innovative Dental Technologies, Inc.
Must be sold by or on the order of a dental professional only.
Use of this devices is prohibited for any other purpose than recommended by the manufacturer.

દ્વારા ઉત્પાદિત

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (5)INNOVATIVE PRODUCT BRANDS™
Innovative Product Brands Inc. TM

IPB, Inc

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (6)

FAQ

Are there any contraindications for using the ReLeaf Restorative system?

No known contraindications have been identified.

What precautions should be taken while using the product?

Risk of allergic reactions cannot be excluded for those predisposed to allergies.

How should loose connections be addressed?

If you find the coupler becoming loose, it may be necessary to replace the O-ring.

દસ્તાવેજો / સંસાધનો

ReLeaf RESTORATIVE Basic Setup Kit [પીડીએફ] સૂચના માર્ગદર્શિકા
RESTORATIVE Basic Setup Kit, RESTORATIVE, Basic Setup Kit, Setup Kit, Kit